ECSS Particle Monitoring — Tape-Lift Sampling (space-systems/ecss/q7050-tape-lift-sampling)
Use when the task is the tape-lift part of the surface clause of ECSS-Q-ST-70-50C: lifting the particles resident on a hardware surface with an adhesive tape, counting them under magnification, and reading the surface against its stated cleanliness level.
Domain quick reference
- A lift measures the surface, not the air. It returns everything the surface is carrying, whatever put it there — fallout, handling, machining residue from before the part ever entered a cleanroom.
- A lift is never complete. The tape removes a fraction of what is resident, and that fraction depends on the finish, so the raw count understates the surface until the recovery efficiency divides it.
- The recovery correction has a credibility limit. At low efficiency the correction is contributing more to the reported number than the count is, and the right response is a better method rather than a larger multiplier.
- A blank lift taken on a clean reference coupon carries the tape's own inclusions and the counting background. Without it those are reported as particles found on the hardware.
- A cleanliness level is a cumulative distribution, not a set of independent bands. Each entry means particles at or above that size, so the counts are accumulated from the largest band downwards before anything is graded.
- A level is stated over a reference area, and a lift almost never covers it. Scaling the counts by the area ratio is what makes the two comparable, and it extrapolates: a small patch multiplies both the particles and the luck.
- Lifting more patches is the honest way to raise sampled area. Two lifts of the same size halve the scaling factor, which is a real improvement in the result rather than a change to the arithmetic.
- The surface is graded on its worst band. Averaging the bands lets a clean fine fraction carry a large-particle failure, and it is the large particles a level exists to exclude.
Workflow
- Validate the lift: sampled area, recovery efficiency on that finish, and the counted size bands.
- Subtract the blank lift band by band, refusing a blank counted on bands the sample was not counted on, and flooring a band at zero rather than returning a negative count.
- Divide each band by the recovery efficiency to recover the resident population from the lifted one.
- Accumulate the bands from the largest size downwards into the cumulative form a cleanliness level is stated in.
- Scale the cumulative counts from the total lifted area — patch area times the number of lifts — onto the reference area of the level.
- Grade each band against its own allowance, absorbing an equality at the allowance with a tolerance rather than relaxing it, and refuse a band the level states no allowance for.
- Report the governing band by utilisation, the failed bands, and the findings: a missing blank, a lifted area below the minimum, and a recovery efficiency low enough that the correction dominates.
Pitfalls
- Reporting the raw lifted count. It is a fraction of what is on the surface, and the fraction moves with the finish, so an uncorrected count is not comparable between two parts, let alone against a level.
- Rescuing a poor recovery with a large multiplier. Below roughly half recovery the number reported is mostly the correction, and the measurement it rests on is the weaker part of the product.
- Skipping the blank. Tape carries its own inclusions and the counting method has its own background, and both are indistinguishable from surface particles once they are in the tally.
- Grading the bands as independent counts. A level is cumulative, so a band read on its own understates the population at or above that size and passes surfaces the level was written to reject.
- Scaling from a tiny patch. The area ratio multiplies the particles and the sampling variance together, so a small lift produces a confident-looking number built on whether one large particle was inside the patch.
- Averaging the bands. The large-particle bands are the reason a level exists, and an average lets an unusually clean fine fraction cover a failure in exactly the band that matters.
- Lifting from a bare optical surface. The method removes particles by contact and can take coating with them; a co-located witness coupon carries the sample instead.
Behavior contract (gate 3)
The lift validation, blank subtraction with its band matching, recovery efficiency correction, cumulative accumulation, reference-area scaling, per-band grading with its at-allowance tolerance, the governing band and the lift findings are exercised by the gate 3 contract test: scripts/test_q7050_tape_lift_sampling.py against scripts/q7050_tape_lift_sampling_logic.py (stdlib unittest, offline). Run: python3 scripts/test_q7050_tape_lift_sampling.py
Compliance
- ECSS standards are freely downloadable (ESA); cite the source and paraphrase per standards-map.yaml.