Managing Cardiac Devices
Interprets pacemaker and ICD interrogation data with programming optimization documentation.
Why This Skill Exists
Over 1.2 million cardiac implantable electronic devices (CIEDs) are implanted annually worldwide. Device interrogation data contains critical information about arrhythmia events, lead function, battery status, and therapy delivery that directly impacts patient safety. Failure to recognize lead malfunction, inappropriate ICD shocks, or suboptimal pacing parameters can result in syncope, cardiac arrest, or unnecessary therapies that degrade quality of life.
The HRS (Heart Rhythm Society) Expert Consensus on CIED management mandates systematic interrogation review, remote monitoring enrollment, and structured follow-up. This skill ensures every device check is documented completely with actionable findings.
Checkpoint A: Pre-Draft Intake (Mandatory)
- What device type is implanted — single-chamber pacemaker, dual-chamber pacemaker, ICD, CRT-D, CRT-P, or leadless pacemaker? (default: "Device type not specified")
- What is the device manufacturer and model? (default: "Manufacturer unknown")
- What was the indication for implantation? (default: "Indication not documented")
- When was the device implanted or last generator change? (default: "Implant date unknown")
- Is this a routine interrogation, symptom-driven check, or remote monitoring alert? (default: "Routine follow-up")
- Has the patient reported any symptoms — palpitations, syncope, shocks, dizziness? (default: "No symptoms reported")
- Is there a prior interrogation available for comparison? (default: "No prior report available")
- Is the patient on antiarrhythmic medications? (default: "Medication list not provided")
Documents to Request
- Complete device interrogation printout (all pages including EGMs)
- Prior interrogation report for comparison
- Current device programming summary
- Remote monitoring transmissions (last 3 months)
- 12-lead ECG with and without magnet application
- Echocardiogram (LVEF for ICD/CRT assessment)
- Current medication list (antiarrhythmics, anticoagulants)
- Operative report from implant or last revision
Step 1: Battery and Lead Parameter Assessment
Battery Status Evaluation:
| Status |
Definition |
Action |
| BOL (beginning of life) |
Full battery voltage |
Routine follow-up |
| MOL (middle of life) |
Adequate remaining battery |
Routine follow-up |
| ERI (elective replacement indicator) |
Battery nearing depletion |
Schedule generator replacement within 3 months |
| EOL/EOS (end of life/service) |
Critical battery depletion |
Urgent generator replacement; device may revert to VVI |
Lead Parameter Thresholds:
| Parameter |
Atrial Lead |
RV Lead |
LV Lead (CRT) |
| Pacing threshold |
< 1.5 V @ 0.4 ms |
< 1.5 V @ 0.4 ms |
< 3.5 V @ 0.5 ms |
| Sensing (P/R wave) |
> 1.5 mV |
> 5.0 mV |
N/A (unipolar) |
| Impedance |
200–1500 Ω |
200–1500 Ω |
200–1500 Ω |
Lead Malfunction Red Flags:
- Threshold > 5 V or sudden rise > 1 V from baseline → lead dislodgement, micro-fracture, or fibrosis
- Impedance < 200 Ω → insulation breach
- Impedance > 2000 Ω → lead fracture
- Sensing drop > 50% → lead dislodgement or tissue changes
Step 2: Pacing Function Assessment
Pacing Mode Nomenclature (NBG Code):
- Position I: chamber paced (A, V, D)
- Position II: chamber sensed (A, V, D)
- Position III: response to sensing (I = inhibited, T = triggered, D = dual)
- Position IV: rate modulation (R)
- Position V: multisite pacing (for CRT)
Common Pacing Modes:
| Mode |
Typical Indication |
| AAI(R) |
Sinus node dysfunction with intact AV conduction |
| VVI(R) |
Atrial fibrillation with slow ventricular response |
| DDD(R) |
Sinus node dysfunction or AV block with sinus rhythm |
| DDI(R) |
Paroxysmal AF with AV block (avoids tracking) |
Assess:
- Percent atrial pacing (% AP) — high AP in DDD may suggest sinus node disease
- Percent ventricular pacing (% VP) — minimize RV pacing if LVEF preserved (algorithms: MVP, AAISafeR, RYTHMIQ)
- Rate response settings — appropriateness for patient's activity level
- Mode switch episodes — frequency and appropriateness (AF/flutter detection)
Step 3: Arrhythmia Episode Review (ICD/CRT-D)
Episode Classification:
| Zone |
Typical Rate (bpm) |
Typical Therapy |
| VT Monitor |
130–170 |
Monitor only, no therapy |
| VT-1 |
170–200 |
ATP × 3, then cardioversion |
| VT-2/VF |
> 200 |
ATP during charging, then defibrillation |
| SVT discriminators |
Variable |
Withhold therapy if SVT confirmed |
For Each Stored Episode, Document:
- Date and time
- Heart rate at onset and during episode
- Rhythm classification by device (VT, VF, SVT, noise)
- Whether device classification was correct (review EGMs)
- Therapy delivered (ATP, cardioversion, defibrillation) and result
- Whether therapy was appropriate or inappropriate
Inappropriate Shock Causes:
- Atrial fibrillation with rapid ventricular response
- Sinus tachycardia (exercise)
- Lead noise / oversensing (T-wave oversensing, lead fracture artifact)
- Electromagnetic interference
- Double-counting (RBBB counting R and R' as separate events)
Step 4: CRT Optimization (CRT-D/CRT-P)
Biventricular Pacing Assessment:
- Target: ≥ 98% biventricular pacing
- Common reasons for low BiV%: frequent PVCs, AF with fast rate, loss of LV capture, inappropriate AV delay
Optimization Parameters:
- AV delay optimization: guided by echo (Ritter method or iterative LVOT VTI)
- VV delay optimization: simultaneous vs. LV-first (typically 0–40 ms LV pre-excitation)
- Monitor for LV lead dislodgement: rising threshold, phrenic nerve stimulation, loss of capture
CRT Response Assessment:
- Responder: ≥ 5% improvement in LVEF and/or ≥ 1 NYHA class improvement
- Super-responder: LVEF normalizes (> 50%)
- Non-responder: check BiV%, lead position (lateral/posterolateral preferred), AV/VV timing, AF burden, scar burden
Step 5: Documentation and Follow-Up Plan
Structured Interrogation Report Must Include:
- Device type, manufacturer, model, serial number
- Battery status with estimated longevity
- Lead parameters: threshold, sensing, impedance (each lead)
- Pacing percentages (atrial and ventricular)
- Arrhythmia episode summary with EGM review conclusions
- Therapy summary (ATP events, shocks — appropriate vs. inappropriate)
- Programming changes made and rationale
- Remote monitoring status confirmation
- Next scheduled follow-up
Follow-Up Intervals (HRS Recommendations):
- In-person: every 6–12 months
- Remote monitoring: every 3 months (or per manufacturer protocol)
- Post-implant: 2–12 weeks wound check + first interrogation
- ERI/EOL: monthly until generator change
- After any therapy or alert: urgent evaluation
Checkpoint B: Post-Draft Alignment (Mandatory)
- Are battery status and estimated longevity documented?
- Are all lead parameters within acceptable ranges or flagged?
- Have all stored arrhythmia episodes been reviewed and classified as appropriate/inappropriate?
- Is the biventricular pacing percentage reported for CRT devices?
- Are programming changes documented with clinical rationale?
Quality Audit
Guidelines
- Always review stored electrograms (EGMs) for every episode — device rhythm classification is not always correct.
- High ventricular pacing percentage (> 40%) in patients with preserved LVEF should prompt evaluation for pacing-induced cardiomyopathy and possible CRT upgrade.
- For inappropriate shocks, identify the cause before reprogramming — T-wave oversensing, lead fracture, and SVT each require different solutions.
- In CRT non-responders, ensure BiV pacing is ≥ 98% before concluding CRT failure — AF, PVCs, and loss of LV capture are correctable causes.
- MRI-conditional devices require specific protocols — confirm device and all leads are MRI-conditional, program to MRI mode, and interrogate immediately post-scan.
- Patients with ICDs who are at end of life should be offered deactivation of shock therapies as part of goals-of-care discussions — document clearly.
- Remote monitoring reduces time to clinical decision for arrhythmias, lead failures, and battery alerts — enrollment is a Class I recommendation.
- Never discharge a patient with an ERI/EOL device without a scheduled generator replacement — document the surgical plan and date.
1---
2name: managing-cardiac-devices
3language: en
4description: Interprets pacemaker and ICD interrogation data with programming optimization documentation. Use when reviewing device interrogations, managing pacemaker settings, or documenting ICD therapies.
5tags:
6 - management
7 - cardiology
8metadata:
9 source: https://skillmd.com/skills/casemark/managing-cardiac-devices
10 author: casemark
11 practice_areas:
12 - Cardiology
13 - Interventional Cardiology
14 - Electrophysiology
15 document_types:
16 - Management Report
17 skill_modes:
18 - Management
19 - Coordination
20---
21
22# Managing Cardiac Devices
23
24Interprets pacemaker and ICD interrogation data with programming optimization documentation.
25
26## Why This Skill Exists
27
28Over 1.2 million cardiac implantable electronic devices (CIEDs) are implanted annually worldwide. Device interrogation data contains critical information about arrhythmia events, lead function, battery status, and therapy delivery that directly impacts patient safety. Failure to recognize lead malfunction, inappropriate ICD shocks, or suboptimal pacing parameters can result in syncope, cardiac arrest, or unnecessary therapies that degrade quality of life.
29
30The HRS (Heart Rhythm Society) Expert Consensus on CIED management mandates systematic interrogation review, remote monitoring enrollment, and structured follow-up. This skill ensures every device check is documented completely with actionable findings.
31
32---
33
34## Checkpoint A: Pre-Draft Intake (Mandatory)
35
361. What device type is implanted — single-chamber pacemaker, dual-chamber pacemaker, ICD, CRT-D, CRT-P, or leadless pacemaker? (default: "Device type not specified")
372. What is the device manufacturer and model? (default: "Manufacturer unknown")
383. What was the indication for implantation? (default: "Indication not documented")
394. When was the device implanted or last generator change? (default: "Implant date unknown")
405. Is this a routine interrogation, symptom-driven check, or remote monitoring alert? (default: "Routine follow-up")
416. Has the patient reported any symptoms — palpitations, syncope, shocks, dizziness? (default: "No symptoms reported")
427. Is there a prior interrogation available for comparison? (default: "No prior report available")
438. Is the patient on antiarrhythmic medications? (default: "Medication list not provided")
44
45### Documents to Request
46
47- Complete device interrogation printout (all pages including EGMs)
48- Prior interrogation report for comparison
49- Current device programming summary
50- Remote monitoring transmissions (last 3 months)
51- 12-lead ECG with and without magnet application
52- Echocardiogram (LVEF for ICD/CRT assessment)
53- Current medication list (antiarrhythmics, anticoagulants)
54- Operative report from implant or last revision
55
56---
57
58## Step 1: Battery and Lead Parameter Assessment
59
60**Battery Status Evaluation:**
61
62| Status | Definition | Action |
63|--------|-----------|--------|
64| BOL (beginning of life) | Full battery voltage | Routine follow-up |
65| MOL (middle of life) | Adequate remaining battery | Routine follow-up |
66| ERI (elective replacement indicator) | Battery nearing depletion | Schedule generator replacement within 3 months |
67| EOL/EOS (end of life/service) | Critical battery depletion | Urgent generator replacement; device may revert to VVI |
68
69**Lead Parameter Thresholds:**
70
71| Parameter | Atrial Lead | RV Lead | LV Lead (CRT) |
72|-----------|------------|---------|---------------|
73| Pacing threshold | < 1.5 V @ 0.4 ms | < 1.5 V @ 0.4 ms | < 3.5 V @ 0.5 ms |
74| Sensing (P/R wave) | > 1.5 mV | > 5.0 mV | N/A (unipolar) |
75| Impedance | 200–1500 Ω | 200–1500 Ω | 200–1500 Ω |
76
77**Lead Malfunction Red Flags:**
78- Threshold > 5 V or sudden rise > 1 V from baseline → lead dislodgement, micro-fracture, or fibrosis
79- Impedance < 200 Ω → insulation breach
80- Impedance > 2000 Ω → lead fracture
81- Sensing drop > 50% → lead dislodgement or tissue changes
82
83---
84
85## Step 2: Pacing Function Assessment
86
87**Pacing Mode Nomenclature (NBG Code):**
88- Position I: chamber paced (A, V, D)
89- Position II: chamber sensed (A, V, D)
90- Position III: response to sensing (I = inhibited, T = triggered, D = dual)
91- Position IV: rate modulation (R)
92- Position V: multisite pacing (for CRT)
93
94**Common Pacing Modes:**
95
96| Mode | Typical Indication |
97|------|-------------------|
98| AAI(R) | Sinus node dysfunction with intact AV conduction |
99| VVI(R) | Atrial fibrillation with slow ventricular response |
100| DDD(R) | Sinus node dysfunction or AV block with sinus rhythm |
101| DDI(R) | Paroxysmal AF with AV block (avoids tracking) |
102
103**Assess:**
104- Percent atrial pacing (% AP) — high AP in DDD may suggest sinus node disease
105- Percent ventricular pacing (% VP) — minimize RV pacing if LVEF preserved (algorithms: MVP, AAISafeR, RYTHMIQ)
106- Rate response settings — appropriateness for patient's activity level
107- Mode switch episodes — frequency and appropriateness (AF/flutter detection)
108
109---
110
111## Step 3: Arrhythmia Episode Review (ICD/CRT-D)
112
113**Episode Classification:**
114
115| Zone | Typical Rate (bpm) | Typical Therapy |
116|------|-------------------|-----------------|
117| VT Monitor | 130–170 | Monitor only, no therapy |
118| VT-1 | 170–200 | ATP × 3, then cardioversion |
119| VT-2/VF | > 200 | ATP during charging, then defibrillation |
120| SVT discriminators | Variable | Withhold therapy if SVT confirmed |
121
122**For Each Stored Episode, Document:**
1231. Date and time
1242. Heart rate at onset and during episode
1253. Rhythm classification by device (VT, VF, SVT, noise)
1264. Whether device classification was correct (review EGMs)
1275. Therapy delivered (ATP, cardioversion, defibrillation) and result
1286. Whether therapy was appropriate or inappropriate
129
130**Inappropriate Shock Causes:**
131- Atrial fibrillation with rapid ventricular response
132- Sinus tachycardia (exercise)
133- Lead noise / oversensing (T-wave oversensing, lead fracture artifact)
134- Electromagnetic interference
135- Double-counting (RBBB counting R and R' as separate events)
136
137---
138
139## Step 4: CRT Optimization (CRT-D/CRT-P)
140
141**Biventricular Pacing Assessment:**
142- Target: ≥ 98% biventricular pacing
143- Common reasons for low BiV%: frequent PVCs, AF with fast rate, loss of LV capture, inappropriate AV delay
144
145**Optimization Parameters:**
146- AV delay optimization: guided by echo (Ritter method or iterative LVOT VTI)
147- VV delay optimization: simultaneous vs. LV-first (typically 0–40 ms LV pre-excitation)
148- Monitor for LV lead dislodgement: rising threshold, phrenic nerve stimulation, loss of capture
149
150**CRT Response Assessment:**
151- Responder: ≥ 5% improvement in LVEF and/or ≥ 1 NYHA class improvement
152- Super-responder: LVEF normalizes (> 50%)
153- Non-responder: check BiV%, lead position (lateral/posterolateral preferred), AV/VV timing, AF burden, scar burden
154
155---
156
157## Step 5: Documentation and Follow-Up Plan
158
159**Structured Interrogation Report Must Include:**
1601. Device type, manufacturer, model, serial number
1612. Battery status with estimated longevity
1623. Lead parameters: threshold, sensing, impedance (each lead)
1634. Pacing percentages (atrial and ventricular)
1645. Arrhythmia episode summary with EGM review conclusions
1656. Therapy summary (ATP events, shocks — appropriate vs. inappropriate)
1667. Programming changes made and rationale
1678. Remote monitoring status confirmation
1689. Next scheduled follow-up
169
170**Follow-Up Intervals (HRS Recommendations):**
171- In-person: every 6–12 months
172- Remote monitoring: every 3 months (or per manufacturer protocol)
173- Post-implant: 2–12 weeks wound check + first interrogation
174- ERI/EOL: monthly until generator change
175- After any therapy or alert: urgent evaluation
176
177---
178
179## Checkpoint B: Post-Draft Alignment (Mandatory)
180
1811. Are battery status and estimated longevity documented?
1822. Are all lead parameters within acceptable ranges or flagged?
1833. Have all stored arrhythmia episodes been reviewed and classified as appropriate/inappropriate?
1844. Is the biventricular pacing percentage reported for CRT devices?
1855. Are programming changes documented with clinical rationale?
186
187---
188
189## Quality Audit
190
191- [ ] Device type, manufacturer, model documented
192- [ ] Battery voltage and estimated longevity stated
193- [ ] All lead thresholds, sensing amplitudes, and impedances recorded
194- [ ] Lead parameters compared to prior interrogation
195- [ ] Pacing mode and percentages documented
196- [ ] Rate response settings reviewed for appropriateness
197- [ ] All arrhythmia episodes reviewed with EGM analysis
198- [ ] Inappropriate therapies identified and programming adjusted
199- [ ] CRT biventricular pacing percentage reported (if applicable)
200- [ ] CRT optimization status documented (AV/VV timing)
201- [ ] Remote monitoring enrollment confirmed
202- [ ] Programming changes listed with rationale
203- [ ] Follow-up interval and plan documented
204- [ ] Patient symptoms correlated with device data
205
206---
207
208## Guidelines
209
2101. Always review stored electrograms (EGMs) for every episode — device rhythm classification is not always correct.
2112. High ventricular pacing percentage (> 40%) in patients with preserved LVEF should prompt evaluation for pacing-induced cardiomyopathy and possible CRT upgrade.
2123. For inappropriate shocks, identify the cause before reprogramming — T-wave oversensing, lead fracture, and SVT each require different solutions.
2134. In CRT non-responders, ensure BiV pacing is ≥ 98% before concluding CRT failure — AF, PVCs, and loss of LV capture are correctable causes.
2145. MRI-conditional devices require specific protocols — confirm device and all leads are MRI-conditional, program to MRI mode, and interrogate immediately post-scan.
2156. Patients with ICDs who are at end of life should be offered deactivation of shock therapies as part of goals-of-care discussions — document clearly.
2167. Remote monitoring reduces time to clinical decision for arrhythmias, lead failures, and battery alerts — enrollment is a Class I recommendation.
2178. Never discharge a patient with an ERI/EOL device without a scheduled generator replacement — document the surgical plan and date.