1---2name: hls-identify-fda-presub-issues3description: Reviews a pre-submission package for a novel medical device to identify issues in product classification, biocompatibility categorization, software concern classification, electromagnetic compatibility testing gaps, investigational study requirements, predicate selection, imaging-compatibility claims, drug-use instructions, and clinical study design deficiencies.4---56# Skill: Identify Issues in FDA Pre-Submission Package for Novel Thrombectomy Device78## 2. Failure modes the skill is correcting910- Product classification, combination-product pathway, and intended-use framing are accepted too quickly, causing the submission to be routed under the wrong regulatory theory.11- Biocompatibility scope is keyed to the wrong contact type or duration, so the proposed test set does not match the device’s actual patient-contact profile.12- Software concern level is understated relative to the device’s clinical function, especially where software failure can affect treatment decisions, device control, or procedural safety.13- Electromagnetic compatibility testing is omitted or treated as optional even though a catheter-based hospital-use device may face standard EMC expectations.14- Imaging-compatibility claims are stated without the bench, safety, or performance support needed to sustain them.15- Clinical investigation planning fails to account for investigational-device requirements, oversight obligations, adverse-event reporting, follow-up adequacy, and exclusion criteria.16- Predicate selection is too narrow, relying on a single comparator when intended use or technological characteristics suggest a split or secondary predicate discussion is needed.17- Instructions for use do not adequately address drug co-use, handling, delivery, compatibility, or user steps where the device is intended to work with therapeutic agents.18- Issues are described narratively without prioritization, regulatory grounding, or concrete recommendations.1920## 3. Legal frameworks / domain conventions that apply2122- Product classification and pathway: assess whether the device is a stand-alone device, part of a combination product, or otherwise subject to a different center-led review framework; if drug or biologic elements are present, confirm that the primary mode of action analysis is documented under the applicable combination-product consultation framework.23- Biocompatibility: classify patient contact by tissue/contact type and duration, then match that classification to the appropriate biocompatibility test battery under accepted device-biological evaluation principles.24- Software concern: evaluate the clinical consequence of software failure under the FDA software documentation framework and do not accept a low concern classification if the software can materially affect diagnosis, therapy, or procedural control.25- EMC: confirm that the test plan addresses applicable electromagnetic compatibility expectations for hospital and clinical environments under the relevant FDA and consensus-standard framework.26- Imaging compatibility: if the device claims imaging compatibility, verify that the claim is tied to a test plan that actually supports the claimed performance and safety under relevant imaging-related standards and FDA expectations.27- Investigational use: if the draft contemplates a clinical study of a significant-risk device or a noncleared use, assess whether an investigational authorization and associated IDE framework apply before initiation.28- Predicate selection: for substantial-equivalence analysis, the comparator should share intended use and enough technological similarity to support the claimed comparison; where the comparison is mixed, a secondary or split-predicate discussion may be necessary.29- Drug-use instructions: where the device is intended to be used with drugs, the IFU should address handling, delivery, compatibility, preparation, and operator directions consistent with the relevant drug-device use context.30- Clinical study design: review the protocol for independent oversight, adverse-event reporting, monitoring, duration of follow-up, and exclusion criteria under the applicable human-subjects and device-investigation framework.3132## 4. Analytical scaffolds33341. Regulatory pathway triage: identify whether the package is describing a device-only submission, a combination-product scenario, or a device with drug-adjacent claims; then test whether the pathway analysis is complete and internally consistent.352. Contact-profile check: identify the patient-contact surface, tissue type, and duration category used in the package; compare it to the intended clinical use and flag any mismatch in the biocompatibility plan.363. Software-risk check: identify the stated software concern level, the software functions implicated, and the clinical consequence of failure; if the classification appears understated, identify the more defensible level and the reason.374. EMC and environmental testing check: identify the proposed bench and environmental testing set; verify whether EMC is expressly included and whether the planned conditions match the intended use environment.385. Imaging-claim check: if the package asserts imaging compatibility or imaging visibility, test whether the claim is supported by specific performance and safety testing rather than labeling alone.396. Predicate analysis: identify the proposed predicate, the intended use, and the core technological differences; assess whether a single predicate is sufficient or whether a secondary or split-predicate explanation would better support the comparison.407. Investigational-study check: identify whether a clinical study is proposed, whether the device and use scenario appear significant-risk, and whether the package addresses the investigational authorization path before study start.418. Drug-use instruction check: if drugs or therapeutic agents are part of the use case, assess whether the IFU and user training materials address preparation, compatibility, delivery, and failure-handling steps.429. Clinical protocol adequacy check: assess oversight, safety monitoring, adverse-event reporting, stopping criteria, follow-up length, and exclusion criteria for fit with the device’s risk profile and study objectives.4310. Prioritization rule: rank issues by whether they threaten the submission pathway, the legitimacy of clinical evidence, or the credibility of a core performance claim.4445## 5. Vertical / structural / temporal relationships4647- Separate threshold issues from downstream implementation issues: pathway/classification defects come first, because they can invalidate the rest of the package.48- Treat testing gaps as upstream of claim support: if a claim lacks the testing needed to sustain it, the claim issue should be linked to the missing test plan, not discussed in isolation.49- When clinical evidence depends on regulatory status, address investigational authorization before protocol sufficiency.50- If one deficiency affects multiple sections of the package, note the interaction once and then anchor the issue to the section where the omission is most visible.51- Where the package spans device description, nonclinical testing, labeling, and protocol materials, analyze each layer against the same intended-use statement to expose internal inconsistency.52- If the document set suggests more than one comparator, pathway, or study scenario, enumerate the alternatives explicitly before evaluating them.5354## 6. Output structure conventions5556- Produce a prioritized issue memorandum suitable for a regulatory review audience.57- Start with a brief executive summary that states the overall risk picture and a concise issue tally by severity using a stated ordinal scale such as Critical / High / Medium / Low.58- Define the severity scale once, then apply it uniformly to every issue entry.59- For each issue, use an industry-conventional issue format with:60 - document section or source location61 - issue description62 - governing regulatory or standards basis63 - severity64 - recommended fix65- Make each issue actionable: identify the problem, state why it matters under the governing authority, and specify the practical correction needed.66- When the package’s documents cross-reference one another, note the interaction so the reader can see whether one omission undermines multiple sections.67- End with a Recommended Actions block that assigns each next step to a role and gives a timing anchor tied to the submission or study milestone.68- Use clear legal and regulatory citations for every proposition relied on; name the controlling rule, guidance, part, or framework rather than stating conclusions without authority.69- Keep the memorandum focused on issues, not a restatement of the whole package.